Preview

Ttttttttttttttttt

Good Essays
Open Document
Open Document
769 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Ttttttttttttttttt
Pore Pressure Transducer (PPT) Calibration Procedure By Javier Ubilla Rensselaer Polytechnic Institute

-

The main components of the PPT are the Head which contains the Diaphragm with 4 strain gauges attached in full bridge configuration. This Diaphragm is protected with a Porous Stone. The sensor is connected to the Data Acquisition System (DAQ) through a Transition Cable and then a Thin Cable. In order to facilitate the saturation of the PPT, especially in tests where viscous fluid is used, at RPI the Porous Stone of the sensor is replaced with filter paper.

-

Estimate the maximum expected pore pressure to be measured during the test and select the appropriate range for the PPT (10, 50, 100 or 200 PSI). Also, according to the location of the sensor determine to use a PPT with a thick cable (Model GE Druck PDCR 81) or with a thin cable (Model GE Druck PDCR 81-3478). For more information about this miniature http://www.gesensing.com/druckproducts/ pressure transducer visit

-

-

Connect the PPT to the National Instruments Data Acquisition System (DAQ), using the Modulus SCXI-1520 and the Accessory SCXI-1314.

-

Input an Initial Calibration Factor (ICF) of 1.0 and NULL the sensor. After the NULL is completed monitor the sensor, check that the sensor reading is stable and with a low level of noise. Blow in the sensor and make sure that the sensor responds to the increase in pressure. Insert the transition cable in the appropriate rubber plug as shown in the figure. Two rubber plugs are available which will perfectly fit the PPT with thick or thin cable.

-

-

Insert the head of the PPT in the triaxial chamber and seal the chamber with the rubber plug as shown in the picture.

-

Turn on the GE Druck Portable Pressure Calibrator model DPI 610/615. This calibration device is being used to produce a stable pressure inside the sealed triaxial chamber. The calibration device has a screen which displays the pressure inside the chamber

You May Also Find These Documents Helpful

  • Better Essays

    Gas Stoich Honors

    • 910 Words
    • 4 Pages

    8. Holding your finger over the hole in the rubber stopper, invert the tube into the 400-mL beaker of water. Clamp the gas collection tube in place on the ring stand so that the stoppered end of the tube is under the water but not touching the bottom of the beaker.…

    • 910 Words
    • 4 Pages
    Better Essays
  • Powerful Essays

    C311

    • 489 Words
    • 4 Pages

    The purpose of the pipet calibration is to determine the accuracy and precision of 10ml of water at room temperature by using a 10ml volumetric pipet. Also, analyzing the analytical balance and the density of water from literature reference.…

    • 489 Words
    • 4 Pages
    Powerful Essays
  • Powerful Essays

    1. Half-fill the test tube with distilled water. Set the tube into the 24-well plate. (The well plate will function as a test tube holder.)…

    • 1713 Words
    • 7 Pages
    Powerful Essays
  • Good Essays

    Lab Report Help

    • 1320 Words
    • 6 Pages

    5. Connect a rubber stopper to the pressure probe and later to the flask with valve closed so that absolutely no hydrogen gas will escape.…

    • 1320 Words
    • 6 Pages
    Good Essays
  • Better Essays

    analysis of alum lab

    • 1065 Words
    • 5 Pages

    2. Tie the capillary tube to the thermometer using a rubber band with it's opening…

    • 1065 Words
    • 5 Pages
    Better Essays
  • Good Essays

    Catalase Lab Report

    • 970 Words
    • 4 Pages

    7. Plug in the Vernier Gas Pressure Sensor onto the Go Link (See Figure 2)…

    • 970 Words
    • 4 Pages
    Good Essays
  • Good Essays

    Ap Biology Enzyme

    • 425 Words
    • 2 Pages

    3) The experimental system of this lab is, first to set up the logger pro pressure measuring device to the computer. After the device is set up completely, set up the logger pro software that will collect the data for this lab. After both of these steps are complete, get a clean test tube with 1.5 mL of H2O, 1.5mL of H2O2 and then add drops of enzyme needed. Collect data by plugging the black “plug” into the test tube and press “start” on the logger pro software. Continue collecting data with variety of different conditions.…

    • 425 Words
    • 2 Pages
    Good Essays
  • Powerful Essays

    Unit two Biology

    • 7492 Words
    • 30 Pages

    A simple respirometer. 13 attached graph paper or ruler 4 At measured time intervals record the position of the inner meniscus in the tube using the attached graph paper or ruler. The volume of gas inside the apparatus should fall steadily as oxygen is used up. 5 Plot a graph of position of meniscus against time and draw a line of best fit.…

    • 7492 Words
    • 30 Pages
    Powerful Essays
  • Powerful Essays

    A lab report at uni

    • 1481 Words
    • 6 Pages

    Where Ps is the starting pressure (before pressurising the vessel), Pi is the intermediate pressure (stabilized once air pump is switched off) and Pf is the final pressure (stabilized after valve opening). You will need to specify the absolute values at each pressure. Note that the equipment will measure the gauge pressure relative to atmospheric pressure. Therefore you will need to add the atmospheric pressure (take this value to be 100 kPa). Ensure you use a steady stable value for each pressure value. Plotting a graph of your data or analysing the data table should help you identify these values. You might consider taking an average value over a short period of time.…

    • 1481 Words
    • 6 Pages
    Powerful Essays
  • Good Essays

    Butane Lab

    • 284 Words
    • 2 Pages

    6. Weigh your dried lighter before the procedure and after the procedure and record the butane gas lost…

    • 284 Words
    • 2 Pages
    Good Essays
  • Good Essays

    6. Rinse both pipets and if any liquid spills rinse once more or until cleaned.…

    • 1072 Words
    • 5 Pages
    Good Essays
  • Better Essays

    Lab 1 Pendulum

    • 1893 Words
    • 8 Pages

    Fill one wide-neck pipet bulb completely with water, invert and quickly place it on top of the test tube 's gas delivery tube as shown in the picture at right.…

    • 1893 Words
    • 8 Pages
    Better Essays
  • Powerful Essays

    Vo2 Max Test Lab Report

    • 3736 Words
    • 15 Pages

    The steps for this is firstly remove the calibration strip from the plastic bag, put the calibration strip into the strip inlet in the test meter, a beep will go off and 888 will appear on the screen. After 2 seconds another beep will go off and function number will on the screen, make that the number that appears matches the number printed on the box and finally put the calibration strip back in the plastic bag. For the rest of the equipment set up it is the same as the VO2 Max Test. In the VO2 Max Test it is explained how to approach this test by doing a PAR Q with the client, measuring the height and weight. Also the equipment is the exact same as the VO2 max test as calibration needs to be done on the mass flow sensor, gas analysers, ECG leads and the mouth piece and nose clips are used in the blood lactate test. Next write the information for the test into the computer which is the warm up time and the workload that is being used. Before the test begins let the client rest on the bike for a minute to get comfortable with the equipment and also to take the clients blood lactate level at rest. After the minute rest the warm up will begin and then the client will go into the exercise test where the workload will be added to the bike. The workload will increase through the test and every 2 minutes in the exercise test the…

    • 3736 Words
    • 15 Pages
    Powerful Essays
  • Powerful Essays

    Properties of Gases

    • 1568 Words
    • 7 Pages

    b. Take the rubber stopper with the pipet hole and gas tubing in the top and cap the test tube tightly. Take some tissue paper, a couple mm thick, and use it to help level your test tube in the well plate. Watch what happens.…

    • 1568 Words
    • 7 Pages
    Powerful Essays
  • Powerful Essays

    venturi meter

    • 914 Words
    • 4 Pages

    determined. The apparatus used on this experiment also has a number of straight tube manometers…

    • 914 Words
    • 4 Pages
    Powerful Essays