Preview

Scientific Measurement and Error

Satisfactory Essays
Open Document
Open Document
662 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Scientific Measurement and Error
EXPERIMENT #1: SCIENTIFIC MEASUREMENT AND ERROR

By:
Date: 8/26/13

Objectives: The goals of the experiment are to learn how to use different types of scientific glassware and to understand the significance of precision and accuracy in scientific measurement.

Procedure: No modifications were made during the experiment.

Data: Data tables for the experiment are provided on attached sheet.

Calculations:
Table I -

Table III –

Table IV –

Conclusion: By performing the experiment I was able to meet the two objections provided. During the first portion of the procedure, an analytical balance (#2) was used to determine the weight of a given bottle, the bottle’s lid, and the combination of the two. After taking four measurements of each, I then calculated the means and standard deviations. For the data recorded in table I the values computed for the standard deviations were very small (.0001 g - .0002 g), so the measurements given by the analytical balance are very precise. After the means were calculated, I compared the weight measured of the lidded bottle to the weight of the same lidded bottle on four different balances. Since the largest gap between all of the results was .0005 g, it’s clear that an analytical balance is an extremely accurate instrument for measuring weight, assuming there are no systematic errors present at the time.
The data recorded in table II was collected by examining the imprints on different types and sizes of volumetric glassware that are shown to determine the calibration temperatures, the volume graduations and whether the glassware was calibrated “to contain” (TC) or “to deliver” (TD) a certain volume of fluid. After studying the different types of glassware, I can determine which is the most suitable for use when performing an experiment.
In the third part of the experiment; 100 mL of deionized water was measured in a graduated cylinder by adding 10 mL increments, then the

You May Also Find These Documents Helpful

  • Good Essays

    6.03 Calorimetry Lab

    • 380 Words
    • 2 Pages

    Class average for a 80 and 120 drop count was recorded to be 0.041+/- 0.005 mL/drop. If a graduated cylinder is not available there is another way of adding more of the solution. The disposable pipet can transfer 0.5mL of the solution to the chemical reaction. When comparing a 10mL graduated cylinder, a 100mL beaker, and a 50mL graduated cylinder, the most precision peace of glass wear would be the 10 mL graduated cylinder with the true volume of water (calculated), class average, and uncertainties being 9.7896 +/- 0.05mL. The 50mL graduated cylinder comes in second place with a true volume of water(calculated), class average, and uncertainties being 50.0001 +/- 0.22mL. The third peace of glassware coming in as the least accurate would be the 100 mL beaker with a true volume of water(calculated), class average, and uncertainties being 59.9999 +/- 0.82mL.…

    • 380 Words
    • 2 Pages
    Good Essays
  • Good Essays

    Ap Chemistry Lab Report

    • 678 Words
    • 3 Pages

    A 125 mL flask was obtained. The square of aluminum was fashioned over the flask by laying the foil over the mouth and folding the sides down. A pin was then obtained and was used to poke a tiny hole in the center of the aluminum cap.…

    • 678 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Stoichiometry Lab Report

    • 1941 Words
    • 8 Pages

    3. After the sample has cooled, add a few drops of deionized water. What happens and what can be concluded?…

    • 1941 Words
    • 8 Pages
    Good Essays
  • Good Essays

    6.03 Calorimetry Lab

    • 992 Words
    • 4 Pages

    Using the buret to read the volume of water is a source of random error because it is marked to the nearest 0.1mL. Therefore, our data could be above or below the actual volume of water in the buret because we took our reading to the nearest 0.01mL, which increases our uncertainty level. Along with taking our volume readings to the nearest 0.01mL, a random error also arose when we took the reading of the temperature of the water. The thermometer only read to the nearest tens places, therefore our readings went to only the tenths place which could have been measured inaccurately giving rise to more uncertainty and random error within the lab. We were able to limit our uncertainty level when taking the masses of the beaker with and without water, by using the analytical balances which read to the nearest ±0.0001g.…

    • 992 Words
    • 4 Pages
    Good Essays
  • Satisfactory Essays

    Priple Beam Balance

    • 213 Words
    • 1 Page

    A graduated cylinder was used for measuring the volumes of the liquid. When measuring the volumes be sure to be eye level with the meniscus. Any other position when looksing at the container will cause an error in the experiment.…

    • 213 Words
    • 1 Page
    Satisfactory Essays
  • Good Essays

    Basic Lab Measurements

    • 465 Words
    • 2 Pages

    This lab experiment displays the accuracy of scientific measuring equipment. This experiment also helps to familiarize the students with these instruments.…

    • 465 Words
    • 2 Pages
    Good Essays
  • Satisfactory Essays

    Chem Lab 2 Report

    • 942 Words
    • 9 Pages

    Procedure: In the experiment, I use devices such as a ruler, beakers, flasks, a scale and cylinders to measure the values of each object and liquid.…

    • 942 Words
    • 9 Pages
    Satisfactory Essays
  • Good Essays

    Volumetric glassware, such as a conical flask, pipette or the burette might have accumulated a systematic error. For example; the burette is not precise when determining the endpoint. It is a random error, but it can also be a systemic error due to the calibration of the glassware used in this experiment. A systematic error occurred during this experiment was the inaccuracy of the measuring scale. This is because the same measuring scale was used in each trial to determine the amount of the antacid tablet. However, the problem was that there was never one significant value mostly because there is no stable reading. To solve this problem, use a more accurate measuring scale to reduce the change in…

    • 1380 Words
    • 6 Pages
    Good Essays
  • Good Essays

    Synthesis Lab

    • 435 Words
    • 2 Pages

    At the beginning of the lab, a micro test tube with and without glass wool was weighed on an analytical balance. A watch glass, crucible, and lid were also weighed. The masses of all the materials needed for the lab were recorded in the data tables of the laboratory notebook. An unknown vial was obtained from the teacher and set to zero on the balance. About 0.6 grams of the unknown was measured into the micro test tube. The analytical balance was set to zero again and then the test tube with the unknown was measured and recorded.…

    • 435 Words
    • 2 Pages
    Good Essays
  • Good Essays

    Procedure: Using distilled water, premeasured containers and objects determine displacement of fluids and density of objects. Use ice and heat measure temperatures in Celsius, Fahrenheit and Kelvin.…

    • 1061 Words
    • 5 Pages
    Good Essays
  • Good Essays

    Which piece of glassware was the most accurate? Why might that be true?  25ml Erlenmeyer flask, this may have been contributed by the shape of the glass and also the way it is calibrated. ©2014,CarolinaBiologicalSupplyCompany 3 MEASUREMENT AND UNCERTAINTY General Lab Questions 1. Consider an experiment to determine the average rate of reaction of an antacid in water.…

    • 589 Words
    • 3 Pages
    Good Essays
  • Powerful Essays

    Finding Density Lab

    • 1151 Words
    • 5 Pages

    been previously done. After the mass of water was measured, the beaker and graduated cylinder were…

    • 1151 Words
    • 5 Pages
    Powerful Essays
  • Satisfactory Essays

    Yes Man

    • 405 Words
    • 2 Pages

    c) (3.15 x 10-1 – 9.7 x 10-3) * [(4.7 x 10-2)/(2.33 x 10-3)] = 1.5 x 10-2…

    • 405 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    The experiments were carried out using a short-wide (SW) glass and a tall-skinny (TS) glass. Orange juice is used as the experiment medium. On visual inspection, these glasses look different in size and shape. Volume of the glasses is given by the relation,…

    • 825 Words
    • 4 Pages
    Good Essays
  • Powerful Essays

    Nt1310 Unit 6 Lab Conclusion

    • 2742 Words
    • 11 Pages

    The mass of a cylinder was determined using a digital balance. These results are more questionable since the device may not have been properly calibrated. A more refined approach would be to use a newly calibrated balance of even greater…

    • 2742 Words
    • 11 Pages
    Powerful Essays