Preview

BIocem Lab: Measuring Spectra of Methyl Orange and Bromophenol Blue

Powerful Essays
Open Document
Open Document
875 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
BIocem Lab: Measuring Spectra of Methyl Orange and Bromophenol Blue
I. Abstract
The purpose of this practical was to measure the amount of a chemical substance present in a sample. Primarily, the aim of Experiment 1 was to measure the absorption spectrum of a particular coloured substance (in this case Bromophenol Blue and Methyl Orange) at varying wavelengths of light. The goal of this experiment was to measure the given amounts of a given chemical substance that was present in a given sample of Methyl Orange and in blood. We measured the optical density at wavelengths 400 to 700 in intervals of 20 mu. For Experiment 2, the process of the experiment focuses on the substance and records its absorption levels at different concentrations. The purpose of this experiment was for each student to test their blood glucose level and determine if they had low blood glucose, which indicate hypoglycemia, or high blood glucose, which indicate hyperglycemia.
Introduction
II. Introduction
In the first part of the experiment we utilized absorbance measurements using a UV-visible light spectrophotometer to determine the concentrations of a substance. A spectrophotometer consist of a monochromator for the selection of wavelength, a sample holder referred to as a cuvette, a recorder, and a light detector called a phototube. We then went on to measure
The absorption levels (A) that are obtained when samples of any given substance exposed to a source of visible light are measured by: The length (l) of the light path (i.e. the distance over which the light had to travel), and the concentration (c) of that substance. This, essentially, is the Beer-Lambert law. Therefore, a molecule can only absorb a specific amount of visible light because of the size of its container (the length from one side to the other) and the concentration.
Absorption levels vary across compounds. To measure the absorbance, a reference sample has to be used along with the actual dye sample being tested. The intensity of the light passing through the reference cell is

You May Also Find These Documents Helpful

  • Better Essays

    Bio Spectroscopy Lab

    • 1273 Words
    • 6 Pages

    This investigation used spectroscopy to evaluate light absorption in different solutions. A spectrophotometer was used in the lab to determine these values. A spectrophotometer is an apparatus used to “measure the absorption of radiation in the visible and UV regions of the spectrum and allows precise at a particular wave length” (Jones et al., 2007). The amount of light absorbed by a substance is directly in relation to the concentration of the solute and also the wavelength moving through the solute (Jones et al., 2007). This is commonly referred to as Beer’s Law and can be expressed as A= εl [C]. Beer’s Law equation measures the absorbency of light, making it an effective measure as spectrophotometers give exact values for absorbency (Jones et al., 2007).…

    • 1273 Words
    • 6 Pages
    Better Essays
  • Satisfactory Essays

    In order to create a standard curve we created 5 dilutions of solutions containing red dye #40. Concentrations of .666 ug/mL, 1.667ug/mL, 3.333ug/mL, 13.333ug/mL, and 20.000 were measured in a spectrophotometer set to a wavelength of 504 nm to find the amount of light they absorbed. Using Microsoft Excel, we entered the absorbance values as the Y coordinates and the concentrations as the X coordinates to make a line of best fit and that was our standard curve.…

    • 342 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    CHEM 1252 Lab Report 5

    • 1825 Words
    • 21 Pages

    For the first part of the experiment, 4.0 mL of the 0.2 M NaOH solution were poured into a small test tube. The test tube was wiped and placed into the sample holder of the spectrophotometer and the transmittance was set to 100.0 for calibration. The test tube was removed from the sample holder and 1 drop of phenolphthalein was added and inverted several times to mix. With the Spec-20 set to absorbance mode, the test tube was quickly placed into the sample holder and the absorbance was recorded every 5 seconds for 360 seconds. For the second part of the experiment, the same calibration steps were repeated and then, 2.0 mL of the 0.2 M NaOH and 2.0 mL of the 0.2 M NaCl were added to a small test tube. After this step, 1 drop of phenolphthalein was added and inverted several times to mix. With the Spec-20 set to absorbance mode, the test tube was quickly placed into the sample holder and the absorbance was recorded every 5 seconds for 360 seconds. After each part of the experiment, the test tube was removed from the sample holder and its contents were placed in the waste.…

    • 1825 Words
    • 21 Pages
    Good Essays
  • Satisfactory Essays

    Chemistry 116 lab review

    • 289 Words
    • 2 Pages

    The first step is to calibrate the colorimeter with0.20 M Fe(NO3)3and set the absorbance at 470 nm since it is known to keep an acidic solution throughout the entirety of the experiment. It was important to do this right at the beginning of the lab since the zeroed value of the acid was the calibration number for all of the other solutions. A total of seven solutions with different dilutions were used throughout the lab to conduct the equilibrium constant. The first step was adding 5 mL of 0.200 M Fe(NO3)3to each of the 5 test tubes. Once this was done, 0.00200 M NCS was added to the test tubes, each receiving a different amount; test tube one received 1 mL NCS-and with each test tube the amount of NCS-would increase by 1 mL, test tube 5 received 5 mL of NCS. . The next step was adding HNO3 to each test tube in different volumes; Test tube one received 10 mL of HNO3 and with each test tube the amount of HNO3 decreased by 1 mL, test tube five had no HNO3 added to it. The addition of these solutions formed five test tubes of different dilutions, but of equal volume, 10 mL each. After all of the previous trials had been completed the final step was to take each test tube and pour it into a different cuvette and measure the absorbance for each. Once the initial concentration was calculated of Fe3+, NCS and FeNCS2+ in molarity. The absorbency values were recorded and used to calculate the formation constant, K f The reference table containing volumes used in each solution is provided below…

    • 289 Words
    • 2 Pages
    Satisfactory Essays
  • Powerful Essays

    Penny Lab Report

    • 2257 Words
    • 10 Pages

    Introduction: In the spectrophotometric analysis of a penny lab, we will be conducting three experiments. The purpose of the first experiment is to find the maximum absorbance on the spectrometer to obtain the optimum wavelength. Once we have gathered that information, we will use the results in the other two experiments. The purpose of…

    • 2257 Words
    • 10 Pages
    Powerful Essays
  • Good Essays

    Food Dyes Lab

    • 671 Words
    • 3 Pages

    The initial step to determine the concentration of the dyes previously found was to prepare a set of serial dilutions at 20%/40%/60%/80%/100% dye concentration. To prepare the 20% dilution 1mL of red 40/blue 1 was combined with 4mL of DI water. This method was used to produce the remaining dilutions with 2mL/3mL, 3mL/2mL, 4mL/1mL, and 5mL/0mL (mL dye / mL DI water). Each of the previous dilutions was tested in the spectrometer and data was recorded. Using the data obtained two Beer Law plots were plotted for both dyes and trendlines were added. The slope of the trendline was found and used to calculate the actual (mol/L) and percent concentrations…

    • 671 Words
    • 3 Pages
    Good Essays
  • Satisfactory Essays

    Dye Analysis Lab Report

    • 978 Words
    • 4 Pages

    2) Based on maximum absorbance of the dye tested, select the appropriate wavelength on spectrometer.…

    • 978 Words
    • 4 Pages
    Satisfactory Essays
  • Satisfactory Essays

    Lab Writeup

    • 237 Words
    • 2 Pages

    Purpose: To measure the amount of pigment in a solution of 40% ethanol and deionized water both containing pieces of an onion with a spectrometer.…

    • 237 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    In order to complete the objective, the group used a Carolina Digital Spectrophotometer, unit M 201-1, which measures the quantity of light that solutions absorb. The spectrophotometer consists of a light source that shines through a prism which then shines through a…

    • 1159 Words
    • 5 Pages
    Good Essays
  • Satisfactory Essays

    Unit1 Forensics

    • 363 Words
    • 2 Pages

    5.) Spectrophotometry is a method to measure how much a chemical substance absorbs light by measuring the intensity of light as a beam of light passes through sample solution. It is most useful when identifying whether or not there is a drug present.…

    • 363 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Spinach Lab

    • 767 Words
    • 4 Pages

    To prepare the solution, 4 mL of 80% acetone was pipetted into a test tube, and five drops of the top layer of the pigment extract from the first experiment was combined with the acetone. The test tube was inverted until thoroughly mixed. Then a test tube containing 4 mL of 80% acetone was created to blank the spectrophotometer at 380 nm. The absorbance of the tube with the extract and acetone solution was then tested for its absorbency. These tests were repeated with the same solution at 40 nm intervals, until the wavelength reached 740 nm, re-blanking the machine between each test or change in wavelength. All the absorbencies were recorded, and the areas with the highest absorbance were retested at 5 nm intervals. The results were then plotted on Excel to show the relationship between certain wavelengths and the absorbance of the…

    • 767 Words
    • 4 Pages
    Good Essays
  • Better Essays

    Copper In Brass Lab

    • 1053 Words
    • 5 Pages

    One test tube was then filled half full with the same solution of nitric acid and wiped clean of finger prints and smudges. This sample was used as a blank to set the zero absorbance for the spectrophotometer. The direction of the test tube was marked and noted for future use. The student then selected a New Absorbance vs. Concentration experiment in SpectroPro. This was done in order to calibrate the spectrophotometer. The wavelength setting was set on the spectrophotometer at 620nm and the machine was set absorbance, listed as unit (A). The student set the blank to zero absorbance and finished the calibration of the spectrophotometer. The solution in the test tube was then discarded and the test tube was set…

    • 1053 Words
    • 5 Pages
    Better Essays
  • Good Essays

    Kinetic Reaction Lab

    • 719 Words
    • 3 Pages

    The third part of the experiment was performed using the aforementioned proportions of the solutions. 4.8 mL of blue food coloring solution and 3.2 mL of NaOCl were mixed and placed in the Spectronic 20. The transmittance of the mixture was recorded at one minute intervals. By using Eq. 1, these values were converted to obtain the absorbance of the blue food coloring. The absorbance was then used to determine the concentration of the blue food coloring by using the following…

    • 719 Words
    • 3 Pages
    Good Essays
  • Best Essays

    Spectroscopy is the study of light. A spectrophotometer is a machine used to determine the absorbance of light at any given wavelength. It does this by using a source of white light through a prism, which gives multiple wavelengths that can be individually focused (Ayyagari and Nigam, 2007). Substances are put into cuvettes that are glass or quartz containers that light can easily travel through. The light that is being focused travels through the substance gets absorbed by the substance and is reflected back and read by galvanometer which had the ability to detect electric currents (Verma, R). The absorbance reading is then given, absorbance is usually between 0.0 – 2.0, any higher than 2.0 may mean not enough light is getting through to the galvanometer (Bhowmik and Bose, 2011). When using the spectrophotometer it is necessary to use a control or blank to zero or tare the machine in between every new wavelength or concentration, this control is water (Ayyagari and Nigam, 2007). The correlation between the numbers acquired through spectroscopy can be seen using the Beer- Lambert Law. The Beer- Lambert law states that the amount of light absorbed at a certain wavelength is proportional to the concentration of the absorbing substance (Fankhauser, 2007).…

    • 2210 Words
    • 9 Pages
    Best Essays
  • Best Essays

    The UV/Vis spectrophotometer work by passing the light by strong lamp through a monochromator, which split the light into its component colours using a gratings. Then, the desired wavelength of light beam is direct through the sample compartment which the cuvette is placed by using the mirror and slit. At other side of the compartment, there is a detector which can determine the amount of light beam pass through the solution in the cuvette, thus the absorbance of the solution in the cuvette can be determined. In addition, the absorbance of light is determined by three factors which is the nature of the solution, concentration of the solution, and the distance of light travel through the solution (path length) which stated in the Beer-Lambert Law.…

    • 2163 Words
    • 9 Pages
    Best Essays