Preview

Standardization Of Edta Lab Report

Satisfactory Essays
Open Document
Open Document
657 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Standardization Of Edta Lab Report
Vu Ngo
% MgO Determination
September 10, 2012

Procedure:
Reference to lab manual
Data:
Mass of Na2EDTA∙2H2O, g | .9289 |
Standardization of EDTA Solution | Trial 1 | Trial 2 | Final buret reading, EDTA (mL) | 18.5 | 36.7 | Initial buret reading, EDTA(mL) | 0.5 | 18.5 |
Determination of % MgO of Unknown Unknown Number | 4J | | Sample 1 | Sample 2 | Sample 3 | Mass of sample (g) | 0.2135 | 0.2132 | 0.2139 | Final buret reading, EDTA (mL) | 73.5 | 74.2 | 74.2 | Initial buret reading, EDTA(mL) | 0.5 | 0 | 0 |
Measurement of water blank Final buret reading, EDTA(mL) | 14.5 | Initial buret reading, EDTA(mL) | 14.2 |
Sample Calculation:
Given:
Molar mass of Na2EDTA*2H2O, g/mol | 372.25 | Expected M of EDTA
…show more content…
The measurement using the the buret have a +/- .1mL error. These types of measurement error are based on the sensitivity of the instrument, during the measurement. These errors could cause the result to deviate from the actual answer.
During the experiment, using titration to find the end point. Because of the end point was unknown, a single drop could determine its end point. There might be an extra drop that caused the data to vary as both parts of the experiment have determined. The first trial of both the experiment, the number is different, because of the unsure number of drops. Using the result from the first trial, the result of the second trial was more accurate. As proven from the experiment of determination of percent of MgO of the unknown. The sample 2 and sample 3 testing results in similarity that differ from the sample one. The sample used 73.0 ml of the EDTA solution, while both samples 2 and 3 used 74.2 ml of EDTA solution. This will cause a tip in the result toward a lower number used to find the end point. This error is not a major factor since it is only of my 1.2 ml of the actual result. The percent of all three result of MgO is different under 1 percent different. Thus, a very reasonable result in all three of the

You May Also Find These Documents Helpful

  • Good Essays

    Chem Pre-Lab

    • 635 Words
    • 3 Pages

    In Figure 2, the lower meniscus lies below the 3 mL line and 4 minor gradations that signify 0.1 values. Estimating the uncertain digit to be 0.04 mL, the reading we record for this particular buret is 3.44 mL.…

    • 635 Words
    • 3 Pages
    Good Essays
  • Good Essays

    6.03 Titration Lab

    • 573 Words
    • 3 Pages

    4.) Fill the buret with NaOH. Record the initial volume of the buret in your data table.…

    • 573 Words
    • 3 Pages
    Good Essays
  • Powerful Essays

    Diprotic Acid Lab Report

    • 1283 Words
    • 6 Pages

    This would give inaccurate results due to misreading the burette. If you estimate the readings on the burette, your results will not be precise and the titres many not be similar. Not viewing the meniscus at eye level would affect the titres because cause the titre values to vary due to guessing what the reading is on the meniscus. Poor lighting could occur which would make it difficult to look for subtle colour change, making the experimenter adding too much sodium hydroxide into the food acid solution because they didn’t see the first permanent colour change. Expelling the sodium hydroxide from the burette into the conical flask of food acid too fast, which would affect the titre, due to adding too much or not enough sodium hydroxide into the food acid. Misjudging the end point, which would affect the titre because more or less solution may have entered then the previous time, causing the titres results to vary. Adding too much or too little indicator, can affect your results due to the end point of the experiment occurring faster or slower than the previous…

    • 1283 Words
    • 6 Pages
    Powerful Essays
  • Better Essays

    6.03 Calorimetry Lab

    • 1068 Words
    • 5 Pages

    Besides, the conical flask should be stirred all along the titration to ensure the two solutions can be mixed evenly. Furthermore, repeat the experiment for getting the average readings to ensure the accuracy of the experiment. Lastly, keep the eyes in level with the liquid surface while reading the burette to prevent parallax…

    • 1068 Words
    • 5 Pages
    Better Essays
  • Good Essays

    This could be due to the many assumptions being made in this experiment. We are assuming that there was no cross contamination between Solution A, B and water. Though it is highly likely that someone used the same pipette or measuring cylinder to measure chemicals as all three solutions were clear and colourless as notes in the overall observations. This could affect the content of each of the beakers as it would change the concentrations and/or volume of the solutions which was to be kept constant (refer to Table 1.1) greatly affecting the data obtained.…

    • 1146 Words
    • 5 Pages
    Good Essays
  • Good Essays

    Basic Lab Measurements

    • 465 Words
    • 2 Pages

    This lab experiment displays the accuracy of scientific measuring equipment. This experiment also helps to familiarize the students with these instruments.…

    • 465 Words
    • 2 Pages
    Good Essays
  • Good Essays

    For the measurement below carefully read the ruler to the best degree of accuracy. The reading you make in centimeters should have two decimals. In other words can you estimate the difference between 10.1 cm and 10.2 cm? If so, you would report 10.10 cm or 10.15 cm whichever your eyes see. If you see 10 cm, than you would report 10.00 cm. The mm reading may be estimated to +/- 0.1 mm. Refer to the lab for more information. You can obtain an accurate ruler from the web, if you don’t have one.…

    • 1509 Words
    • 7 Pages
    Good Essays
  • Satisfactory Essays

    Chem 142 Lab 1

    • 786 Words
    • 4 Pages

    This experiment is dealing with the relative accuracy of an individual measurement. Data will be provided to perform the calculations asked for. The purpose of this experiment is to know how to use significant figures and to get formularized with making proper laboratory reports.…

    • 786 Words
    • 4 Pages
    Satisfactory Essays
  • Powerful Essays

    Although this experiment is supports the hypothesis enough to arrive at a conclusion like the one above, it did have some sources of error that might have affected the measurements therefore leading to invalid results. Some of these errors…

    • 1267 Words
    • 6 Pages
    Powerful Essays
  • Good Essays

    The calculated volume measurement is more accurate because Archimedes’ Principle might have a higher percentage of error.…

    • 412 Words
    • 2 Pages
    Good Essays
  • Powerful Essays

    Hypothesis/Prediction: The percent composition by mass of magnesium in magnesium oxide will not change significantly with each group that conducted the experiment. The composition of each substance should stay the same and any differences must be due to some error.…

    • 1424 Words
    • 5 Pages
    Powerful Essays
  • Powerful Essays

    Finding Density Lab

    • 1151 Words
    • 5 Pages

    was done according to the first but the volume was different, instead of 10 ml, 20 ml was measured.…

    • 1151 Words
    • 5 Pages
    Powerful Essays
  • Better Essays

    Lab Three

    • 1684 Words
    • 7 Pages

    As important as a measurement is there a knowledge of the uncertainty associated with the. (For instance, when using instrument such as scales and thermometers, there is always a margin of error.) In our lab work it is important to keep such uncertainties in mind and use conversions to measure such measurement against other units of measurement measurements to insure the most accurate data during laboratory experiments. The utilization…

    • 1684 Words
    • 7 Pages
    Better Essays
  • Powerful Essays

    Also, not all z.o.i. were a perfect circle, some were oddly shaped which meant you had to use you best judgement on where to start your measurements. This may have led to some inconstancies of results. In future experiments when measurements are involved, the use of the same instrument could minimize errors.…

    • 939 Words
    • 4 Pages
    Powerful Essays
  • Satisfactory Essays

    EE3114 Lab Report

    • 1405 Words
    • 8 Pages

    Dynamic systems like dc-servomotors, financial systems, logistic models, internet systems and eco-systems can be described by a set of coupled differential equations. Based on this model, one can study the behavior of such a system under various external factors such as initial conditions, variables’ interrelation changes, stead state responses and stability issues. In this experiment, a simple Loika-Volterra type model was chosen to demonstrate these behaviors. The concept of state trajectory, limit cycles and stability will be introduced.…

    • 1405 Words
    • 8 Pages
    Satisfactory Essays