Preview

Specific Heat Capacity

Good Essays
Open Document
Open Document
551 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Specific Heat Capacity
Marybeth Brooks

Physics Lab III

Specific Heat Capacity Lab
May 25, 2011

Abstract The specific heat capacity of various metals can be calculated and compared to accepted literature values. Dropping heated metal samples into a calorimeter filled with water and then measuring the change in the temperature of the system accomplished this. The metal samples were heated in a boiling water bath and were assumed to be at 100 C when they were removed. It was added to a calorimeter and stirrer with a measured mass of room temperature tap water. The temperature was taken through the lid of calorimeter to mitigate heat loss. These temperatures were recorded and used to calculate the specific heat capacity of the metal samples. The specific heat capacity of the aluminum was found to be 1.44 J/(g*C) and the relative error was found to be 0.54. The specific heat capacity of the copper sample was found to be 0.499 J/(g*C) and the relative error was found to be 0.112. The specific heat capacity of the iron was found to be 0.596 J/(g*C) and the relative error was found to be 0.148. The values found for the copper and the iron were slightly higher than the accepted literature value but still rather close. The value found for the aluminum was much higher than the accepted.

Sample Calculations -see attached sheet.

Data Analysis

Table of Data collected to measure specific heat capacity.

| |Aluminum |Copper |Iron |
|Metal mass, g |21.84 |72.04 |63.21 |
|Initial Metal Temp, C |100 |100 |100 |
|Initial Temp of System, C |21.5 |21.1 |20.5 |
|Final Temp of System, C

You May Also Find These Documents Helpful

  • Good Essays

    Mole Formula Lab

    • 723 Words
    • 3 Pages

    In order to determine how much of each separate component a total mass was taken before hand for both the aluminum cup and the alum. These weights were shown to be 4.5g for the aluminum cup and 2.0g for the alum crystals. Once the alum was placed over the heat source it began to turn to a cleared liquid and started to bubble. Three trials were conducted and with a constant time over the heat source each time. (10 mins) The results of each trial are found in Table 1 below:…

    • 723 Words
    • 3 Pages
    Good Essays
  • Good Essays

    PHYS172 Lab14

    • 2023 Words
    • 10 Pages

    Today, you will calculate the specific heat capacity per atom of aluminum and of lead…

    • 2023 Words
    • 10 Pages
    Good Essays
  • Good Essays

    Chetos

    • 971 Words
    • 4 Pages

    The amount of heat needed to increase the temperature of an object exactly 1°C is the heat capacity of that object. The specific heat capacity of a substance is then the amount of heat it takes to raise the temperature of 1g of the substance1°C. Caliometry is the precise measurement of the heat flow into or out of a system for chemical and physical processes. Heat flow is measured in two common units, the calorie and the Joule. 1 calorie= 1 kilocalorie = 1000 calories. A calorimeter is the device used to measure the absorption or release of heat. In this lab, the water in aluminum can “calorimeter” gains the heat lost during the combustion of a Cheeto.…

    • 971 Words
    • 4 Pages
    Good Essays
  • Satisfactory Essays

    chem report

    • 601 Words
    • 3 Pages

    First we chose an unknown metal, in which our unknown was unknown metal 2, the metal was gray and irregular shaped. We setup the calorimeter and got a mass of it just being empty then we got a mass with 30mls of water in it. We found the mass of the water by subtracting the mass of the empty calorimeter which was 49.987g and the mass of the calorimeter with water which was 87.332g and got the mass of just the water 37.345g. The metal was placed in a test tube 5cm high and we got the mass of the empty test tube that was 42.703g and then with the metal in it and it was 72.607g. We heated the 300ml of water to a boil, then placed the metal that was in a tube and waited 10 minutes before we took it out. The temperature of the water before the metal was placed was 20 degrees Celsius, when we put the hot metal into the water and stirred it with the stirring rod it was 29 degrees Celsius, the change of the temperature after the metal was placed was 9 degrees. We ran this experiment twice to compare the results. In the second experiment we found that the mass of the calorimeter and water with 50mls was 100.033g and the empty calorimeter was 50.857g. We subtracted them two and we got the mass of water was 49.176g. The temperature of the water before the metal was placed in was 21 degrees Celsius and then when the metal was heated up for 10 minutes the temperature went up to 28 degrees Celsius. The change in temperature in the second trial was 7 degrees Celsius.…

    • 601 Words
    • 3 Pages
    Satisfactory Essays
  • Satisfactory Essays

    Practice Final

    • 1884 Words
    • 8 Pages

    | A 5.10-g sample of iron is heated from 36.0oC to 75.0oC. The amount of energy required is 89.54 J. The specific heat capacity of this sample of iron is…

    • 1884 Words
    • 8 Pages
    Satisfactory Essays
  • Good Essays

    CHM 130 Lab 4

    • 1761 Words
    • 8 Pages

    1. Calculate the change in temperature for the water caused by the addition of the aluminum by subtracting the initial temperature of the water from the…

    • 1761 Words
    • 8 Pages
    Good Essays
  • Satisfactory Essays

    Chemistry 1 Lab Report

    • 580 Words
    • 3 Pages

    Conclusion: In this experiment we had to find the heat capacity of the calorimeter cup using two trials of hot and cold water. When we obtained the data after 10 minutes of recording 30 second intervals of the calorimeter cup temperature, we created a…

    • 580 Words
    • 3 Pages
    Satisfactory Essays
  • Good Essays

    Specific Heat Lab

    • 688 Words
    • 3 Pages

    5) Quickly transfer the metal to the calorimeter, immediately cover with an insulating lid and stir. Record the final temperature without touching the metal or the sides of the calorimeter (28.0oC)…

    • 688 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Chemistry 17.1 - 17.4

    • 439 Words
    • 2 Pages

    the heat capacity of an object depends on both its mass and its chemical composition…

    • 439 Words
    • 2 Pages
    Good Essays
  • Satisfactory Essays

    Calometry Lab

    • 587 Words
    • 3 Pages

    For #1 theres a specific heat of 0.864 J / (g × °C) and that is closest to the specific heat of aluminum. So, for this experiment, let's call your metal aluminum.…

    • 587 Words
    • 3 Pages
    Satisfactory Essays
  • Good Essays

    This is known as specific heat capacity; it measures the amount of energy needed to raise the temperature of that particular substance by 1 degree Celsius. Specific heat capacity was first discovered in the 18th century by a Scottish scientist by the name of Joseph Black who noticed that different substances of equal masses required distinctive amounts of heat to raise them to similar temperature intervals (Encyclopædia Britannica 2014). The specific heat capacity of water is 4.184 J/g°C. Energy is most commonly measured joules (J) or kilojoules (kJ) during the study of heat transfer between two substances (Tuckerman n.d.); however, there is another measurement for energy which is used to measure the amount of energy in food, and this is referred to as food calories. Food calories are equal to 1000 heat calories and 1 heat calorie is equal to 4.184 joules (Calories in Food – A Factor in Weight Maintenance and Weight Loss n.d.). Consequently, it can be seen that the specific heat of a substance can also be measured in calories. Therefore, the specific heat capacity of water can also be written as 1 cal/g°C. In this lab, the calculations were done using food calories and heat calories in order to determine the caloric content of one marshmallow per…

    • 774 Words
    • 4 Pages
    Good Essays
  • Better Essays

    In this experiment, the specific heat and the density of an unknown metal was determined in order to identify the unknown metal. The average specific heat of the unknown metal was 0.197˚C and was determined using a calorimeter. The density of the unknown metal was 6.57 g/mL and was determined using a cylinder and displacement. Using the specific heat value of the unknown metal and its density, it was determined that the unknown metal was Tin.…

    • 1030 Words
    • 5 Pages
    Better Essays
  • Satisfactory Essays

    Possible sources of error could be that heat was lost to the environment during the heating process, making the metal with the water in the calorimeter not the correct temperature. This would skew the specific heat value. Another source of error would be human error in reading the temperature level because it was hard to get the exact correct value with the instruments we were using. When weighing the metal, if there were any substances on it or the weight plate, this could have given the wrong mass value. If I were to redo this experiment, the ways I would prevent these errors from occurring would be to obtaining higher value equipment that will ensure exact values. To prevent heat being lost, I would place the system in a vacuum to ensure no heat is lost to the outside. And to ensure no human errors, I would thoroughly clean all materials and if possible substances to ensure all values are correct.…

    • 325 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Calorimetry Intro

    • 717 Words
    • 4 Pages

    In this experiment, a styrofoam calorimeter was utilized to obtain the changes in enthalpy of the various reactions. Before being able to do this, however, the heat capacity of the calorimeter, Ccal, must be determined. This is accomplished by the neutralization of a strong acid and strong base wherein the change of enthalpy of the reaction is already known.3 Having found the Ccal, the calculations for the enthalpy changes and the specific heat of a metal can be performed.…

    • 717 Words
    • 4 Pages
    Good Essays
  • Better Essays

    Rdr 1

    • 1041 Words
    • 5 Pages

    Preferably, it is better to know the heat capacity of the calorimeter itself rather than the entire calorimeter system.…

    • 1041 Words
    • 5 Pages
    Better Essays