Absorption Spectra

Powerful Essays
Introduction

At any particular wavelength the absorption of light when visible light passes through a solution depends on two factors: * The length of the light path * The concentration of the colored path
The connection of these two variables is known as the Beer-Lambert Law:

Absorbance (A) = ε c l

Where ε is the absorption coefficient
C is the concentration of the compound
And l is the length of light usually 1cm

When I is constant, this proves a linear relationship between absorbance and concentration. This means it can be rearranged to find the find the optimum concentration for maximum absorbance.

C = A/ ε

Methods:

The two reagents used are Bromophenol blue and Methyly orange.

Experiment 1:

1 Set up the colorimeter at 400nm, measure the absorbance of both reagents against distilled water, this will be used as a blank
(make sure to zero the colorimeter after each interval)

2 Repeat this procedure at 20nm from 420nm to 700nm

3 After getting a range of results make further readings in smaller intervals to precisely locate the wavelength for λmax

Experiment 2:

1 Using the protocol provided prepare a series of dilutions in testubes for each compound for the range of concentrations from 1 to 10mg/l.

2 Using the λmax for each reagent from Experiement 1 set the colorimeter against distilled water and measure the absorption

3 Repeat for other dilution series

Results:
Experiment 1: Bromophenol Blue | | Further readings to determine λ max | | Wavelength (nm) | Absorbance | Wavelength (nm) | Absorbance | 400 | 0.093 | 595 | 0.98 | 420 | 0.113 | 590 | 0.942 | 440 | 0.035 | 610 | 0.672 | 460 | 0.048 | | | 480 | 0.069 | | | 500 | 0.111 | | | 520 | 0.154 | | | 540 | 0.301 | | | 560 | 0.451 | | | 580 | 0.728 | | | 600 | 0.976 | | | 620 | 0.347 | | | 640 | 0.066 | | | 660 | 0.011 | | | 680 | 0.002 | | | 700 | 0 | | |

Methyl orange | | Further



References: http://www.chemguide.co.uk/analysis/uvvisible/beerlambert.html Accessed: 25/11/12

You May Also Find These Documents Helpful

  • Satisfactory Essays

    Beers law

    • 480 Words
    • 2 Pages

    5. Collect and record the absorbance values for each of the five standards solutions. Print a graph of absorbance vs. concentration.…

    • 480 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Colorimetric Analysis

    • 421 Words
    • 2 Pages

    2. If this analysis was performed using a colorimeter, yellow light would be used to measure the absorbance of solutions. Why is blue light not used?…

    • 421 Words
    • 2 Pages
    Good Essays
  • Satisfactory Essays

    Dye Analysis Lab Report

    • 978 Words
    • 4 Pages

    2) Based on maximum absorbance of the dye tested, select the appropriate wavelength on spectrometer.…

    • 978 Words
    • 4 Pages
    Satisfactory Essays
  • Powerful Essays

    The objectives of this experiment are to understand why and how spectroscopy is used to determine the components and concentrations of a solution, describe various types of spectroscopy, describe the visible and ultraviolet regions of the electromagnetic spectrum, define Beer’s law and define the relationship between absorbance and transmittance. Other learning objectives are to create a Beer’s law plot for a series of samples with known concentrations, collect spectrophotomic data from unknown and known FDC blue dye samples, perform serial dilutions, calculate concentrations, perform linear regression and determine the equation of a best fit line.…

    • 1297 Words
    • 8 Pages
    Powerful Essays
  • Better Essays

    Bio Spectroscopy Lab

    • 1273 Words
    • 6 Pages

    This investigation used spectroscopy to evaluate light absorption in different solutions. A spectrophotometer was used in the lab to determine these values. A spectrophotometer is an apparatus used to “measure the absorption of radiation in the visible and UV regions of the spectrum and allows precise at a particular wave length” (Jones et al., 2007). The amount of light absorbed by a substance is directly in relation to the concentration of the solute and also the wavelength moving through the solute (Jones et al., 2007). This is commonly referred to as Beer’s Law and can be expressed as A= εl [C]. Beer’s Law equation measures the absorbency of light, making it an effective measure as spectrophotometers give exact values for absorbency (Jones et al., 2007).…

    • 1273 Words
    • 6 Pages
    Better Essays
  • Satisfactory Essays

    This lab will teach me how to use a spectrophotometer. The use of the spectrophotometer is to measure the concentration of solute. The solute being measured must be colored and is determined based on the adsorption of light photons on a wavelength. The spectrophotometer uses a beam of light that strikes the diffraction grating that basically forms of prism of light. Then only a specific wavelength of light shines through the spectrophotometer and interacts with the solute. The light that continues past the solute hits the phototube. The spectrophotometer then digitally shows the amount of units that have been absorbed or transmitted. Transmittance is the amount of light that gets through the sample. This is shown as a percent of all the possible light that could’ve gotten through. Absorbance is the opposite of transmittance and the reciprocal of it. This shows how much light got trapped in the solute. In this lab we will use a few different solutions in the spectrophotometer to get a basic feel on how it works. We will record the absorbance as well as do calculations using mean and standard deviation. We will then graph our results and compare them with the class values of the three unknown Methylene blue samples. At the end of the lab the actual concentrations of each unknown will be shared. We will then compare how accurate and precise our results are with the actual.This lab will teach me how to use a spectrophotometer. The use of the spectrophotometer is to measure the concentration of solute. The solute being measured must be colored and is determined based on the adsorption of light photons on a wavelength. The spectrophotometer uses a beam of light that strikes the diffraction grating that basically forms of prism of light. Then only a specific wavelength of light shines through the spectrophotometer and interacts with the solute. The light that continues past the solute hits the phototube. The spectrophotometer…

    • 1420 Words
    • 6 Pages
    Satisfactory Essays
  • Good Essays

    Predictions: Chlorophyll a absorbs best at a wavelength of 400, Chlorophyll b at 450, carotene at 500, and xanthophyll at 450.…

    • 420 Words
    • 2 Pages
    Good Essays
  • Better Essays

    After that take them out, mix the pairs and measure the absorbance the same way as before.…

    • 944 Words
    • 4 Pages
    Better Essays
  • Satisfactory Essays

    Chemistry 116 lab review

    • 289 Words
    • 2 Pages

    The first step is to calibrate the colorimeter with0.20 M Fe(NO3)3and set the absorbance at 470 nm since it is known to keep an acidic solution throughout the entirety of the experiment. It was important to do this right at the beginning of the lab since the zeroed value of the acid was the calibration number for all of the other solutions. A total of seven solutions with different dilutions were used throughout the lab to conduct the equilibrium constant. The first step was adding 5 mL of 0.200 M Fe(NO3)3to each of the 5 test tubes. Once this was done, 0.00200 M NCS was added to the test tubes, each receiving a different amount; test tube one received 1 mL NCS-and with each test tube the amount of NCS-would increase by 1 mL, test tube 5 received 5 mL of NCS. . The next step was adding HNO3 to each test tube in different volumes; Test tube one received 10 mL of HNO3 and with each test tube the amount of HNO3 decreased by 1 mL, test tube five had no HNO3 added to it. The addition of these solutions formed five test tubes of different dilutions, but of equal volume, 10 mL each. After all of the previous trials had been completed the final step was to take each test tube and pour it into a different cuvette and measure the absorbance for each. Once the initial concentration was calculated of Fe3+, NCS and FeNCS2+ in molarity. The absorbency values were recorded and used to calculate the formation constant, K f The reference table containing volumes used in each solution is provided below…

    • 289 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Food Dyes Lab

    • 671 Words
    • 3 Pages

    The initial step to determine the concentration of the dyes previously found was to prepare a set of serial dilutions at 20%/40%/60%/80%/100% dye concentration. To prepare the 20% dilution 1mL of red 40/blue 1 was combined with 4mL of DI water. This method was used to produce the remaining dilutions with 2mL/3mL, 3mL/2mL, 4mL/1mL, and 5mL/0mL (mL dye / mL DI water). Each of the previous dilutions was tested in the spectrometer and data was recorded. Using the data obtained two Beer Law plots were plotted for both dyes and trendlines were added. The slope of the trendline was found and used to calculate the actual (mol/L) and percent concentrations…

    • 671 Words
    • 3 Pages
    Good Essays
  • Powerful Essays

    4. Test the solubility of each substance by mixing small amounts in a separate test tube with 3mL of distilled water in it, then “flick” each tube to mix the contents.…

    • 1321 Words
    • 6 Pages
    Powerful Essays
  • Powerful Essays

    4. Then use the spectrophotometer to measure the #3 abs = 0.334 @ 27 sec. absorbance of each solution.…

    • 568 Words
    • 3 Pages
    Powerful Essays
  • Good Essays

    In this experiment the betalain will be measured by using a colorimeter. A colorimeter is an instrument which compares the amount of light getting through a solution (Betalain) with the amount which can get through a sample of which is clear (distilled water). A reading is then given of which shows the percentage of absorbance. This gives the results for the experiment.…

    • 623 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Enzyme Activity Lab Report

    • 1074 Words
    • 5 Pages

    Test tube one 1was filled with 5.0mL of buffer, 2.0mL of H2O2, and 1.0 mL of guaiacol. Tube 1 will be to calibrate the spectrometer. Tube 2-3 will be filled with Dil 0.25, tube 2, 4, 6 & 8 will contains 2.0 of 10mM H2O2 and 1.0 of 25mM of guaiacol. Tube 3 will contain 4.75 of 0.1 Buffer in mL and 0.25 of enzyme extract in mL. Tube 4-5 will be filed with Dil 0.5, but tube 5 will contain 4.5 of 1.0 Buffer in mL and 0.5 of enzyme extract in mL. Tube 6-7 will be filled with Dil 1.0, for instant tube 7 will contain 4.0 of 0.1 Buffer in mL and 1.0 enzyme extract in mL. Tube 8-9 will be filled with Dil 2.0; in addition tube 9 will contain 3.0 of 0.1 Buffer in mL and 2.0 of enzyme extract in mL. After we were done, we turned the spectrophotometer one and let it warm for about 5 minutes, we set the wavelength to 500nm we set it to a point that we could only read the absorbance until 1.999. We took a cuvette rack and placed about 5 cuvette tube. We took a clean cuvette and transfer all the contents of the tube 1 into the cuvette and wipe the outside with a kimwipe and insert the cuvette into the sample holder. After doing the previous step we turn the 100% transmittance control knob and control it until the reading said…

    • 1074 Words
    • 5 Pages
    Good Essays
  • Good Essays

    In order to complete the objective, the group used a Carolina Digital Spectrophotometer, unit M 201-1, which measures the quantity of light that solutions absorb. The spectrophotometer consists of a light source that shines through a prism which then shines through a…

    • 1159 Words
    • 5 Pages
    Good Essays